Zero Angle Calibration (Version 2.0)
Zero Horizontal Angle Calibration
The HDL32E provides two holes on the bottom side for precision mounting
using dowel pins. In order to reproduce unit to unit reproducibility the zero
angle (for the rotational axis) is calibrated with respect to the dowel pins.
Previously the the zero angle accuracy was limited by manufacturing and
assembly tolerances of several rotational assemblies, while with Version
2.0 the zero angle is calibrated to the dowel pin holes after the assembly
providing much better unit to unit repeatability.
The target is centered to the mounting holes for dowel pins which are
permanently fixed to the floor
Several scans of laser 15 (horizontal) across the target are used to
determine the center position from the edge locations
The accuracy of the calibrated zero angle is about 0.05 degrees
The unit to unit repeatability has additionally been increased with Version
2.0 by not coating the bottom side of the HDL32E sensor, thereby
avoiding inaccuracies for mounting introduced by non-uniform coatings.
Dowel Pin hole Dowel Pin slot
Silver coating has been
removed from bottom side
with version 2.0